The ASRT website will be down for scheduled maintenance from Friday, March 27, at noon to Saturday, March 28, at 12:30 p.m. Mountain time while we make system updates. During this time, you will not be able to access the ASRT site or your ASRT account. We apologize for any inconvenience and appreciate your understanding as we work to improve our services to you.

Dissertation Abstract

Spectral Changes in Electron Beams at Extended Distances Utilizing a Telescoping Intraoperative Radiation Therapy Applicator System

Publication Number: AT1352418

Author: Fadell, Louis

School: Medical College of Ohio at Toledo

Date: 1993

Pages: 53

Subject: Radiation Therapy

Intraoperative Radiation Therapy (IORT), utilizing electron beams, is performed at the Medical College of Ohio at SSD ranging from 100 to 115 cm. This project was designed to investigate possible changes in electron spectra over this range of distances which may impact on the dosimetry as well as shielding of critical organs adjacent to the irradiated site. Electron beam energies of 6, 9, 12, 16, and 20 MeV are evaluated using film dosimetry for changes in Rp, E$/sb[/rm O],$ and Dmax. Results indicate that no significant changes occur in these parameters over the above ranges of distances and energies.